V520 Firmware Release Notes


Table of Contents

WARNING: If you are upgrading from a Beta release of TWB 5.x, you will need to update the firmware twice. Please read Firmware release notes carefully.

The Firmware update procedure has been updated to avoid potential hazards. Please follow the updated instructions detailed in Appendix E of the users' guide.

 

 

1 - Upgrading

1.1 - Firmware

 

1 - Upgrade

1.1 - Firmware

The following procedure should be used to upgrade the firmware on the V520:

1.  Gather and manually write down the configuration of the tester (IP, etc.).
2.  Reset the tester to factory configuration.
3.  Set the tester IP address.
4.  Reboot the tester.
5.  Reconnect with TWB.
6.  Upgrade/downgrade the tester.
7.  Reboot the tester.
8.  Reconnect with TWB.
9.  Reconfigure the tester with the information saved in step 1.